Evaluation of the image quality in computed tomography: different phantoms
DOI:
https://doi.org/10.29384/rbfm.2011.v5.n1.p67-72Abstract
The aim of this paper was to compare the simulators provided by the CT manufactures and Catphan’s Phantom with the American College of Radiology (ACR) computed tomography phantom. The image evaluation followed the protocols established by the manufactures of the phantoms. For slice thickness evaluation, the maximum percentage difference was 9% between the phantoms ACR and Siemens. In CT number accuracy test, the measurements of CT number of water showed a difference of 10 HU between the CT simulators. Comparing the uniformity results, the discrepancy was 11% and 55% for Siemens and Philips respectively in relation to the result obtained with the ACR phantom. The result of low contrast was the same for all phantoms. The MTF50 and MTF10 obtained with Siemens phantom was 4 and 8 pl/mm. For Catphan, 6 and 7 pl/mm. Results demonstrate that the ACR simulator was the most comprehensive and flexible to be used in several scanner models. Some simulators did not present all image quality indicators to perform a complete test.Downloads
Downloads
How to Cite
Issue
Section
License
The submission of original articles to the Brazilian Journal of Medical Physics implies the transfer, by the authors, of the rights of print and digital publication. Copyright for published articles remains with the author, with journal rights on first publication. Authors may only use the same results in other publications by clearly indicating this journal as the original publisher. As we are an open access journal, free use of articles in educational, scientific, non-commercial applications is allowed, as long as the source is cited.
The Brazilian Journal of Medical Physics is under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License (CC BY-NC-ND 4.0).